Articulating, stemmed, antibiotic-impregnated cement spacer using new real components for the treatment of prosthetic knee joint infection
Article History
Abstract
Abstract
Implantation of an antibiotic-impregnated cement spacer following the explantation is a well-accepted treatment strategy for periprosthetic knee joint infection. The indications for static knee spacers are limited owing to increased incidence of joint stiffness, fractures, dislocations with restriction in weight-bearing. The use of real components by autoclaving the explanted components and using them with antibiotic cement to prepare an articulating cement spacer has been described with a good outcome.
Here we present our experience of a modified technique (in three cases) for the preparation of an articulating, stemmed, antibiotic-impregnated cement spacer using new real components. This technique was found to be easily reproducible and allows a functional range of motion and full weight-bearing.
Using real components of TKR, with Ilizarov's components and tension band wire makes a very simple and elegant mobile spacer which has benefits of vastly improved function and good infection control during the intermediate phase and ease of surgery in the second stage.
Keywords
This article was originally published by Reed Elsevier India Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.
1 Introduction
Prosthetic Joint Infection (PJI) is one of the most common causes (20.4%) for revision Total Knee Replacement (TKR). Two-stage revision has been a gold standard treatment modality for chronic and multi-drug resistance organisms.1
Two-stage methods consist of removal of the implant followed by irrigation and debridement followed by implantation of an antibiotic spacer, which can be static or articulating. Static spacers are useful but they cause joint stiffness, fractures, and dislocations.1 Hofmann AA et al., in 1995 first described the use of real components by autoclaving the explanted components and using them with antibiotic cement.2 Since then, many studies have shown similar results using the same, explanted, autoclaved components or using new real components.3,4
Here, we present a modification of the Hofmann technique for the preparation of an articulating, stemmed, antibiotic-impregnated cement spacer using new real components for stage 1 of 2-stage revision total knee replacement.
2 Materials and methods
This study included three patients with a mean age of 62 years. All patients were operated on with primary total knee replacement at 5.3 years on average. Patients presented to our institute with symptoms ranging from knee pain, swelling, redness, instability, and/or discharging sinus at the operated site. All patients were diagnosed with an infected total knee replacement. In joint aspirations, one of the patients had a positive culture for E. Coli and others were culture negative (Table 1). Patients were advised for two-stage revision surgeries according to our protocol (Fig. 1).
| Sr. No. | Age/Sex | Presentation | Pre-op diagnosis | Pre-op culture | Stage 1 culture | Post-spacer antibiotics | Spacer ROM | Spacer duration | Final Surgery | Post stage 2 antibiotics | Complications |
| 1 | 61/F | Joint pain, Swelling, Redness | Infected TKA | E. coli | E. coli | Inj. Ceftriaxone (2g) I.V. every 24 hours + Inj. Linezolid (600mg) I.V. every 12 hours for 21 days | 100–110 | 16 weeks | Hinge knee | Tab. Levofloxacin (500mg) BD + Tab. Linezolid (600mg) BD for 7 days | No |
| 2 | 56/F | Discharge, Joint pain, Instability, Redness | Infected TKA | No growth | S. Aureus | Tab. Levofloxacin (500mg) BD + Tab. Linezolid (600mg) BD for 21 days | 90–100 | 15 weeks | Hinge knee | Tab. Levofloxacin (500mg) BD + Tab. Linezolid (600mg) BD for 7 days | No |
| 3 | 55/M | Discharge, Joint pain, Instability, Redness | Infected TKA | No growth | S. Aureus | Tab. Levofloxacin (500mg) BD + Tab. Linezolid (600mg) BD for 21 days | 90–100 | 10 weeks | Hinge knee | Tab. Levofloxacin (500mg) BD + Tab. Linezolid (600mg) BD for 7 days | No |

2.1 Surgical steps
2.1.1 Stage 1
The knee joint was exposed through a standard approach. Standard radical debridement, removal of implants, and lavage were performed. The removed implants served as a guide for sizing. We selected the appropriate implant size by either upsizing or downsizing the explanted implant. For sizing of the tibial poly, 3–4 mm was added to the thickness of the removed component for the tibial tray thickness. Compensation for ligamentous laxity and bony defects was factored in. The trial reduction was performed with trial components.
A new Posterior Stabilized (PS) femoral component and a new PS tibial poly were used. A 5.5 mm drill bit was used to drill in the center of the post of the tibial poly from the bottom without breaking out at the top. An Ilizarov rod (core diameter 6 mm) of the desired length for the tibial stem was screwed into the poly to make a composite construct. The other end was fixed with two nuts to lock them only if the desired diameter of the rod was 10 mm or higher. If the desired diameter of the rod was less than 10 mm, no locking nuts were used at the other end (Fig. 2).

The femoral stem was prepared of the desired length and diameter depending on the diameter of the last reamer used and its depth. Two (one for femur stem and one for tibial stem) pre-sterile, endotracheal tubes of desired diameters were opened and split open. An Ilizarov rod of the desired length of the femoral stem was taken. A female hinge was used at one end. The other end was fixed with two nuts as described earlier. The rod was placed inside the split open endotracheal tube.
A standard 40 gm. cement pack was taken and mixed. Desired antibiotics, according to previous culture sensitivity reports of the organism identified were added after the initial polymerization had occurred. 4 gm. of sterile Vancomycin powder, 9 million IU of sterile Colistin powder, and 10 ml of sterile Tobramycin liquid (40 mg per ml) were added at this stage per 40 gm. cement pack. After the cement sets, the molds of the endotracheal tubes were removed. The diameter of the stems was re-checked at this stage to confirm that they were not oversized by passing them through an intramedullary nail gauge.
The femoral stem was placed in the canal with the female hinge remaining just outside the canal. A double loop of 1.5 mm tension band wire was passed through the hole in the female hinge and placed over the condyles to have a better cement-steel construct of the distal and posterior augments (Fig. 3). A fresh 40 gm. cement pack was opened and prepared with similar antibiotic addition. The composite tibial component (PS poly with Ilizarov rod and antibiotic cement construct) was cemented in place with a “poor” cementing technique. The femoral component in an appropriate position and rotation was placed over the tension band wire and the hinge with the same “poor” cementing techniques. The poor cementing technique would include: leaving blood to prevent good inter-digitation between bone and cement and pulling out the component a little before final seating to ensure a lack of complete locking. The knee was reduced and put through a range of motion arc and then held until the cement set.

After thorough lavage and hemostasis, the wound was then closed in layers without a drain.
2.2 Post-operative care
Post-operative radiographs were ordered (Figs. 4 and 5). Patients were encouraged to weight-bearing as tolerated as soon as their general condition permitted. Broad-spectrum IV antibiotics were started immediately till the confirmed culture and sensitivity reports were available. All patients were given IV antibiotics during the hospital stay and switched to oral antibiotics for 3 weeks. Serial monitoring of ESR, CRP was done weekly for first two weeks and then, every two weeks until the oral antibiotic therapy. Suture removal was done at 15 days. And the patients were followed up every 15 days until the second stage. At each follow up, patients were examined thoroughly and ESR, CRP was checked. If the ESR, CRP were within normal limits along with clinical and radiological signs of infection control, definitive surgery was planned (Fig. 1).


All three patients had an uneventful postoperative period and were infection-free till the last follow-up.
3 Discussion
Prosthetic joint infection (PJI) is a dreadful complication following total knee replacements. It forms one of the common indications of revisions (20.4%) following total knee replacement and its incidence is increasing.1
In the United States, two-stage revision has been a gold standard for chronic and multi-drug resistance organisms. In India, the incidence of PJI has been found to be around 1.1% in a single institute by Sebastian S et al.5 Majority of them are caused by multidrug-resistant Gram-negative bacteria (64%), Aerobic Gram-negative bacteria (61%) and Methicillin-resistant coagulate negative staphylococcus (45%). Thus, in India, there is a higher need for a two-stage revision TKR for PJI.
Static spacers vs articulating all cement spacers have been compared; where patients with articulating spacers had better ROM, knee score, and less bone loss than patients with static spacers. But, Fink B et al. have shown that these all-cement spacers do abrade within 6 weeks. They suggested radical debridement and total synovectomy with extensive lavage during the second-stage reimplantation to minimize the number of abraded particles.6
Prefabricated molds have shown better results in terms of ROM compared to static spacers but becomes a costlier technique for the patient.1,3 With all-cement spacers, the quantum of cement and antibiotics is fairly high. Acute kidney injury after antibiotic cement spacers has been documented in a fairly high number of patients.7
In 1995, Hoffman AA et al. first described the use of real components by autoclaving the explanted components with antibiotic cement in 26 patients. They showed good results. Since then, many authors have done a similar method and have shown similar results.1,2
Results of two-stage revision between the autoclaved metal-on-cement spacer and static spacer in periprosthetic knee infection have shown similar results in eradicating the infection with both techniques, but ROM after re-implantation was much better with autoclaved components than static spacers.7
We have used new components along with Ilizarov components as a spacer. On the femoral side, along with the real femoral component a threaded rod with a female hinge of Ilizarov in inserted in the medullary canal. It allows us to make the femoral stem of the desired length and diameter. Thus, it provides the desired stability and also provides a greater area of antibiotic elution. It also clears the entire area of interest where the definitive implants would be seated. The threaded rod with nuts at the end aids in the removal of the construct. Using an Ilizarov threaded rod makes the process of spacer removal simple and quick, without leaving any residue inside the canal.
The tension band wire is passed through the female hinge and fixed underneath the cement component. This allows for ensuring a good cement-steel construct over which the component rests. It is also very useful while removal of the spacer as the wires binds the intramedullary stem together with the component.
On the tibial side, a similar threaded rod is put which is fixed with a tibial polyethylene component. In Indian conditions, where the implants are under price control, the cost of an all-polyethylene component is more than twice as much as a regular polyethylene. Thus, using a thicker regular polyethylene and making keel out of bone cement makes it very cost-effective and stable. Also, using cement in the entire keel area would ensure good clearance of infection for use of subsequent metaphysical cones and wedges. Using the threaded rods with nuts for stem would aid in strength and ease of removal. Also, even if, the tibial component poly breaks during removal, the extraction of the threaded rod stem is simple. One can attach a female hinge on top of the threaded rod and back slapped with a regular extractor using the female hinge as the eye of the nail.
Lyons S et al., in 2019 have published their results of using real component knee spacers in 55 patients. Out of these 40% of their patients chose spacer retention at an average of 3.3 years. They concluded that articulating spacers with real components provides excellent infection control and good functional outcome. And the same implant can be retained for an extended period of time similar to a single-stage procedure.3 In 2020, Gililland JM et al. showed their technique of making the articulating spacer with real components. They have used two Steinman pins into the tibial all poly component.9 But, smooth constructs like wires, nails can have complications like cement de-bonding. Cho et al. have shown that using a threaded Ilizarov rod makes the process of spacer removal simple and quick, without leaving any residue inside the canal.10
Spivey JC et al. have done a systemic meta-analysis of 34 articles. They found that metal on polyethylene articulating spacers had a significant increase in joint ROM compared to other types of spacers without difference in infection eradication.8 More recently, Roof MA et al., in 2021 have shown that real components have significantly shorter hospital stays which reduces patient morbidity and mortality. Also, it reduces hospital cost. They concluded that real components shorten operating time, blood loss, improve patient recovery and ease of re-implantation. Thus, making them superior to all-cement spacers.4
In our study, there was improved ROM in all three patients with an average of 100°. At the time of last follow up all patients were symptom-free with control of infections.
Definitive fixation in 2nd stage was uneventful in all three patients.
4 Conclusion
Using real components of TKR, with Ilizarov's components and tension band wire makes a very simple and elegant mobile spacer which has benefits of vastly improved function and good infection control during the intermediate phase and ease of surgery in the second stage.
Financial Support
None.
Patient consent
An informed consent was taken from the patients.
References
- International consensus meeting on musculoskeletal infection: research priorities from the general assembly questions. J Orthop Res. 2018;37(5):997-1006.
- [Google Scholar]
- Treatment of infected total knee arthroplasty using an articulating spacer: 2- to 12-year experience. Clin Orthop Relat Res (430):125-131.
- [Google Scholar]
- Early to midterm results of "low-friction" articulating antibiotic spacers for septic total knee arthroplasty. Arthroplast Today. 2019 Apr 22;5(2):221-225.
- [Google Scholar]
- Comparing the efficacy of articulating spacer constructs for knee periprosthetic joint infection eradication: all-cement vs real-component spacers. J Arthroplasty (21):S0883-S5403.
- [Google Scholar]
- A clinico-microbiological study of prosthetic joint infections in an Indian tertiary care hospital: role of universal 16S rRNA gene polymerase chain reaction and sequencing in diagnosis. Indian J Orthop. 2019 Sep-Oct;53(5):646-654.
- [Google Scholar]
- Articulating spacers used in two-stage revision of infected hip and knee prostheses abrade with time. Clin Orthop Relat Res. 2011 Apr;469(4):1095-1102.
- [Google Scholar]
- Acute kidney injury when treating periprosthetic joint infections after total knee arthroplasties with antibiotic-loaded spacers: incidence, risks, and outcomes. J Bone Joint Surg Am. 2021 May 5;103(9):754-760.
- [Google Scholar]
- Use of articulating spacer technique in revision total knee arthroplasty complicated by sepsis: a systematic meta-analysis. Orthopedics. 2017 Jul 1;40(4):212-220.
- [Google Scholar]
- Balanced, stemmed, and augmented articulating total knee spacer technique. Arthroplast Today. 2020 Oct 1;6(4):981-986.
- [Google Scholar]
- Antibiotic coated hinged threaded rods in the treatment of infected nonunions and intramedullary long bone infections. Injury. 2018 Oct;49(10):1912-1921.
- [Google Scholar]