Design and prototype production of a new external fixator pin that can dampen movement to reduce pin tract inflammation

Author affiliations
Gebze Technical University, Faculty of Engineering, Department of Bioengineering, Kocaeli, 41400, Turkey
Dokuz Eylul University, Institute of Health Science, Department of Biomechanics, Izmir, 35320, Turkey
Dokuz Eylul University, Faculty of Medicine, Department of Orthopaedics and Traumatology, Izmir, 35320, Turkey
Article History
Corresponding author(s)

⁎Corresponding author: Hakan Oflaz. hakanoflaz@gtu.edu.tr

Article tools · PDF & citations

Abstract

Abstract

External fixation is a commonly used method in fixation of bone fractures. In this method, bones are positioned and secured by pins of the external fixator system. Although the system provides a reliable solution for fixing the fracture site, the most common problem in clinical practice of this method is pin site failure due to pin micro motions. This problem leads to the loosening of the pins which results in infection at the pin-bone interface that's called pin tract inflammation. The aim of this study is to reduce pin motions by altering the design of conventional Schanz pin so, a new pin, called Micro-Motion Damping Pin (MDP).

MDP was compared with the Schanz pin in two methods. The first method compared inter-pin motion behaviors on a same external fixator bar that was set on polyethylene rods. The second method compared bending behaviors. A universal axial test machine was used to apply force in the experiments, and a non-contact video extensometer was used to capture inter-pin motions between the first and the third pins on the proximal fragment.

Although, the results showed that there was not a statistically significant differences between two pin designs in reducing the inter-pin motions (p = 0,240), the measured displacement of the MDP was less than the Schanz pin when compared. Displacements were 0,117 ± 0,02mm and 0,142 ± 0,065mm for the MDP and the Schanz pins respectively.

The results point out the MDP absorbed some amount of the inter-pin motion when compared to Schanz pins. Moreover, the MDP's mechanical results were superior in bending tests, and the difference was statistically significant (p = 0,002). In conclusion, the new pin can be used in an external fixation treatment, since its mechanical features are as good as the Schanz pin, however it should be evaluated on human cadaver subjects in future studies.

Abstract

Highlights

•The overall design of the new external fixation pins can improve the anchoring and eliminate the rate of pin loosening.•The new external fixation pin design provides superior dampening of micro motions when compared to conventional Schanz pin.•Under the bending load the new external fixation pin design shows better resistance with respect to conventional Schanz pin.

Keywords

  • External fixator pin
  • Bone fracture
  • Pin-bone interface(s)
  • Pin failure
  • Pin tract inflammation
  • Pin loosening
Disclaimer:
This article was originally published by Reed Elsevier India Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.
1

1 Introduction

External fixation is commonly used in fracture healing and bone lengthening processes. For this reason, mechanical, and biological properties of the external fixation systems are very important.1 In these systems, fixation of pins should be excellent to keep the complete system in stable condition. Therefore, pins play a critical role in stabilization since they transfer the total load from the proximal fragment to the distal fragment via their body by completing the structure with an external fixator bar.2 The mechanical problems of the external fixator pins depend on multiple factors. These factors can be listed as; the number of pins, the pin diameter,3 the pin-bone interface,4 the pin bending stiffness,5 the pin position,6 the pin materials and its coatings.1,7,8 However, the listed factors are not the only ones that effect the stability of the external fixation system.5 For example, pins that are smaller in diameter create micro-motions which lead to early pin loosening and failure, whereas pins that are larger in diameter lead to more stress and bone damage.5

Nevertheless, there are many complications related to the pin-bone interactions such as pin loosening which can cause infection at the connection site.9 In most of the studies that focus on the investigations of pin track infections and delayed healing of the fracture site when using external fixators, it has been observed that sometimes it is required to remove the external fixation system.10 In recent studies, researchers developed novel approaches to the addressed problem of pin loosening caused infection.11–13 These existing problems discourage surgeons from using external fixation techniques. Despite the existing problems, the external fixation is seen as a reliable preference by physicians for treating fractures with defects, infected and non-infected non-unions, delayed fracture union in fracture healing and metaphyseal fractures.14,15

The close relation between pin-track inflammation and loose pins has been mentioned in previous studies.16 One of the causes of bone resorption and subsequent pin loosening is formation of bacteria from the pin-skin interface to the pin-bone interface. Since the pins are the most flexible components of the external fixator systems, pin site failure is the most common problem that is seen faced in the clinical practices of external fixation methods. Thus, it is usually believed that low micro motions cause faster healing and better bone fixation.17,18

The purpose of this study is to design a more stable pin called Micro-Motion Damping Pin (MDP) and compare it with the conventional Schanz pin on polyethylene (PE) rods to examine their mechanical properties. Promising results were recorded that encourage us to carry out this study further on human cadaver subjects.

2

2 Material and methods

2.1

2.1 Pin design

External fixator pins play critical roles in fixation of fractured fragments, so pin designs are extremely important. The conventional Schanz pin has a simple design geometry. It is just like a cylindrical rod that has a threaded frontal part. However, MDP had a different design regarding the Schanz pin. It was comprised of three components as shown in Fig. 1, and it was designed and manufactured by the medical company (HGO Medikal San. Tic. A.S. Izmir, Turkey). The viscoelastic part (A), which was made up of a vulkollan (Ege Asal Metal, Izmir, Turkey), provides low flexibility to the system between the threaded part and a sheath part (B) of the MDP. The last part, which was made up of stainless steel (316L steel), was partially threaded and it was very similar to the design of the Schanz pin. Mechanical features of the MDP may provide more stable fixation by damping the pin motion on a system. The sheath part, which was made up of stainless steel (316L steel), increased the strength of the MDP with a larger diameter design at the middle section. The assembly of the MDP is shown in Fig. 1. In the assembly, the sheath part wraps around the viscoelastic part, and the threaded part is positioned by passing through the hole in the center of the viscoelastic part. When all the parts of the MDP were assembled together, the total length of the MDP was 180 mm which is similar to the conventional Schanz pin.

Design of the Micro-Motion Damping Pin (MDP); Viscoelastic part (A) and sheath part (B) of the MDP.
Fig. 1 Design of the Micro-Motion Damping Pin (MDP); Viscoelastic part (A) and sheath part (B) of the MDP.
2.2

2.2 Experimental setup

MDP was compared with the Schanz pin in experiments with two different measurement methods. The first one was a static axial loading up to 800N to observe the inter pin motion between the first and the third pin on the proximal fragments. The second one was a bending test to observe the maximum loading capacity of the two different pin designs under the same amount of deformation.

2.3

2.3 Polyethylene rods as a fractured bone model

High density polyethylene (PETILEN YY B00552, 0,950–0,956 g/cm3 – PETKIM Holding Company A.S., Turkey) rods (152mm in length and 44mm in diameter) were used to mimic fractured bones as an experimental in-vitro model for both pin designs. These rods were used to form two fragmental parts and align them vertically with the external fixator system. PE rods were cut in the middle to form two fragments to mimic bones undergone osteotomy. Then those two fragments were positioned with the support of a unilateral external fixator system (Fig. 2). The first pin was mounted 30mm below from the top to prevent fatigue. All pins were positioned horizontally in depth of about 40mm into the PE rods. In order to measure the pin stability without any pre-contact between the PE fragments under a load, a 25mm gap was left open between the two fragments. Two different subgroups were formed according to the pin designs and 6 PE experimental models were constituted for each subgroup. Experiments were carried out under the same conditions. Same external fixator bars were used and the functional lengths of both pin designs were the same. Experiments were carried out by a load controlled axial test machine (SHIMADZU AG-100 KNI Model Autography) to apply a static loading (0–800N, 1 cycle and speed = 10mm/min). A non-contact video analysis system (DVE-201 Model Video non-contact extensometer) was used to capture the inter-pin motions between the first and the third pin on the proximal fragment.

Setup of the group of static loading experiments with polyethylene rods; with the conventional AO Schanz pins (a) or the MDP (b). There were two markers on the first and the third pins from the proximal side for the non-contact video extensometer. Under static loading, the micro-motion was recorded for the comparison between the two pin designs.
Fig. 2 Setup of the group of static loading experiments with polyethylene rods; with the conventional AO Schanz pins (a) or the MDP (b). There were two markers on the first and the third pins from the proximal side for the non-contact video extensometer. Under static loading, the micro-motion was recorded for the comparison between the two pin designs.
2.4

2.4 Bending tests

Fig. 3 shows the experimental setup of bending tests. As it can be seen in Fig. 3, the front part of the pins were fixed with a clamp and then ascending force was applied on the rear part of the pins till a certain deformation was reached. The force applied was a gradually increasing force (5mm/min) up to the stroke of the universal axial testing machine reached 50 mm. Force data were recorded for each experiment, and they were compared corresponding to the pin designs. 6 pins were tested for each design.

Setup of the group of bending experiments; with the conventional AO Schanz pins (a) and the MDP (b).
Fig. 3 Setup of the group of bending experiments; with the conventional AO Schanz pins (a) and the MDP (b).
2.5

2.5 Statistical analysis

The statistical analysis of the results was performed via Sigma Plot software version 11.0 for Windows (Systat Software Inc., San Jose, CA, USA). The experimental data included mean ± standard deviation (SD). For a comparative evaluation between the test groups, a Mann-Whitney rank sum test was used. A value of p < 0.01 was considered as statistically significant.

3

3 Results

Two groups were formed according to measurement technique and two subgroups were formed under each main group according to the pin design (Table 1). Firstly, the inter-pin motion was observed on PE specimens. A fractured bone was mimicked using PE rods, which were fixed with the external fixation system by using two different pin designs. Then a static loading up to 800N was performed and the results were compared. 6 measurements were carried out for each subgroup and an average inter pin motion for the MDP between the first and the third pin on the proximal part was measured 0,117 ± 0,02mm, whereas an average inter pin motion for the Schanz pin between the first and the third pin on the proximal part was measured 0,142 ± 0,065mm. The difference was not statistically significant (p = 0,240), however the measured displacement of the MDP was less when compared with conventional AO Schanz pin. This points out the absorption of some amount of the inter-pin motion with respect to Schanz pins. Table 2 shows the data distribution of the current study.

Table 1 Classification of the groups and number of experiments (n) in each sub-group.
SUBGROUPS (Pin Type) MAIN GROUPS
Static Experiments with Polyethylene Rods (n) Bending Evaluation(n)
Conventional Pin 6 6
New Pin (MDP) 6 6
Table 2 Data distribution of all measurements.
Group Test Type Conventional Pin (Mean ± SD) New Pin (MDP)(Mean ± SD) Pa
Experiments with Polyethylene Rods Micro-motion with Extensometer (mm) 0,142 ± 0,065 0,117 ± 0,02 0,240b
Bending Certain Displacement (mm) 31,576 ± 2822 31,729 ± 2849 0,818b
Force (N) at a Certain Displacement 247,667 ± 29,326 330,208 ± 7689 0,002a
The difference is statistically significant respect to Mann-Whitney Rank Sum test.
The difference is not statistically significant respect to Mann-Whitney Rank Sum test.

After the inter-pin motion evaluation, the pins were observed by performing bending tests. The pins were compared under the same circumstances. An axial testing device applied forces up to a certain amount of deformation, and resistive forces were recorded for each pin as shown in Fig. 4. Conventional Schanz pin was capable of resisting 247,667 ± 29,326N at 31,576 ± 2,822mm deformation and the MDP capable of resisting 330,208 ± 7,689N at 31,729 ± 2,849mm deformation as shown in Table 2. Difference between the deformations of two pin designs were not statistically significant (p = 0,818), so it can be assumed that the deformation was same. However, in response to the same deformation values, the pins resisted to different magnitude of forces. As the MDP resisted up to 330,208 ± 7,689N, whereas the Schanz pin resisted up to 247,667 ± 29,326N, and the difference was statistically significant (p = 0,002). These results showed that the MDP can be loaded more at the same deformation.

Plots of the group of bending behaviour; with the conventional AO Schanz pins (a) and the MDP (b). The maximum forces were between 227 N–286 N for the Schanz Pins and 321 N–341 N for the MDP.
Fig. 4 Plots of the group of bending behaviour; with the conventional AO Schanz pins (a) and the MDP (b). The maximum forces were between 227 N–286 N for the Schanz Pins and 321 N–341 N for the MDP.
4

4 Discussion

A key observation on the etiology of the problem in external fixation is that pin loosening and bone resorption can be caused by even a small relative movement at the pin-bone.19–21 As it is known that the stability of the pin-bone fixation is very important, the motion at the site of the pin-bone interface should be reduced to achieve stability. During the fixation state of the fracture site, a small reduction in micromovement at the pin-bone interface may even decrease the loosening.21 In this subject, different designs of pins, insertion methods and coated pins to increase the biocompatibility can be showed as an example like titanium pins,22,23 and enhanced osteointegration of hydroxyapatite-coated pins with slow resorption.1 The mentioned design features may theoretically enhance the rigidity of the pins at the pin bone interface. The experiments with hydroxyapatite-coated pins in animals have confirmed improved pin anchoring and decreased rates of pin loosening.24 Therefore, one of the alternative solutions, based on the clinical results with other orthopedic implants, is the use of hydroxyapatite-coated pins.

There are also different studies about pin designs.25 The basic concepts of these examples are pin diameters, and coaxial half-pin units. For example, pin diameter should be at least 5 mm in the tibial external fixation. Too small diameter pins lead to early pin loosening and failure due to micro-motions. Too large diameter pins can cause stress rising effects, therefore pin core diameter dictates the strength and fatigue resistance of the pins.5 In addition, coaxial half-pin unit can produce and maintain strong preload in the direction of its long axis (axial preload) and can thus achieve a mechanically stable pin bone interface during dynamic loading conditions.26 In addition to all, the stability of the fixation can be provided.

According to a previous study, a prototype pin design produced axial preload, by this way the more stable pin-bone interface was tried to be achieved.26 However, the micro motion could not be absorbed exactly because of the absence of flexibility in the system. Finally, pin problems by micro-motions were not solved efficiently.

Results of a previous study showed that an enhancement of fracture healing was achieved by external application, which limited interfragmentary motions with flexibly fixed fractures.18 By this way the healing sides were stimulated by interfragmentary movements caused by partial or full weight-bearing and low fixation stiffness. For immobilized patients, an externally applied interfragmentary movement (IFM) in the range of 0,5mm may be an appropriate stimulation to enhance fracture healing.18 Furthermore, absence of movement at the pin-bone interface prevents bone resorption and pin loosening.19,20,27,28

According to a previous study, it is generally accepted that small IFMs yield better bone healing results than larger IFMs (>1mm), however the optimal size of IFM within the lmm range remains undetermined.18 Our flexible external fixator pin design contributed to this appropriate healing loading with less pin-bone interaction.

After the review of above-mentioned literatures, some concepts came out to achieve these goals in the reduction of the pin motions for diminishing the inflammation in the treatment of orthopedic surgery. First of all, the viscoelastic material was selected to use in the design of new pin as in one of the previous studies in the literature.18 After the development and testing, the MDP had been formed as an alternative pin. The system depends on the solution of absorbing the micro-motions at a certain rate by the viscoelastic parts. The stability of the fixator systems in orthopedic surgery increases with the large diameter pins, however large diameter pins cause higher stresses and more bone damage. The MDP was designed in different diameters along the length in the current study to be more efficient as mentioned in one of the previous studies in the literature.5

5

5 Conclusion

In conclusion, it was observed that the MDP showed better mechanical behaviors under load than the Schanz pin. Results showed that the MDP was better at absorbing micro pin motions, and it was providing enhanced bending resistance when compared with the conventional Schanz pin. The most common problem in the clinical application of an external fixation, which is the failure at the pin-bone interface and manifests as pin loosening that may lead to pin tract infection or loss of fracture reduction, maybe diminished by MDP.

In the light of the exciting results from this study, in vivo testing on animals would be needed to confirm the results and determine loosening of bone screws on cortical bone during dynamic loading. Therefore, a new future extended study can be planned to improve this study by including the animal experiments. Furthermore, other problems of pins such as pull-out resistance, fatigue, etc. Can be solved by the modification of the MDP.

Ethical statement

This study does not need an Ethical statement.

Patient’s consent statement

This study does not need a Guardian/Patient's consent.

Funding statement

This research did not receive any specific grant from funding agencies in the public, commercial or not-for-profit sectors.

CRediT authorship contribution statement

Muhammed Tarık Çöpoğlu: Conceptualization, Software, Validation, Formal analysis, Investigation, Data curation, Writing – original draft, Writing – review & editing, Visualization. Hakan Oflaz: Conceptualization, Methodology, Investigation, Resources, Data curation, Writing – original draft, Writing – review & editing, Supervision, Project administration. Hasan Havıtçıoğlu: Conceptualization, Resources, Writing – review & editing, Supervision.

References

  1. , , , . Hydroxyapatite coating of threaded pins enhances fixation. The Journal of Bone and Joint Surgery British. 1997;79(3):487-489.
    [Google Scholar]
  2. , , . External fixation: principles and applications. J Am Acad Orthop Surg. 2015;23(11):683-690.
    [Google Scholar]
  3. , , , , . Relationship between rigidity of external fixator and number of pins: computer analysis using finite elements. Revista Brasileira de Ortopedia (English Edition). 2012;47(5):646-650.
    [Google Scholar]
  4. , , , . Bone properties affect loosening of half-pin external fixators at the pin-bone interface. Injury. 2012;43(10):1764-1770.
    [Google Scholar]
  5. , , , . Seven-year outcome following Colles' type distal radial fracture. A comparison of two treatment methods. J Hand Surg. 2003;28 B(5):422-426.
    [Google Scholar]
  6. , , . A safer pin position for external fixation of distal radial fractures. Injury. 2000;31(9):749-750.
    [Google Scholar]
  7. , , , . Role of coatings and materials of external fixation pins on the rates of pin tract infection: a systematic review and meta-analysis. World J Orthoped. 2021;12(11):920.
    [Google Scholar]
  8. , . The influence of hydroxyapatite coating of external fixator pins on pin loosening and pin track infection: a systematic review. Injury. 2010;41(2):128-132.
    [Google Scholar]
  9. , . External fixation using simple pin fixators. Injury. 1992;23(Supplement 4):S1-S54.
    [Google Scholar]
  10. , . Complications of external skeletal fixation. Clin Orthop Relat Res. 1983;180(180):109-116.
    [Google Scholar]
  11. , , , , . Stiffness of a new unilateral external fixator – a biomechanical study. Vibroengineering Procedia. 2020;30:103-108.
    [Google Scholar]
  12. , , , et al . Conventional bicortical pin substitution with a novel unicortical pin in external fixation: a biomechanical study. Injury. 2021;52(7):1673-1678.
    [Google Scholar]
  13. , , , , , , . A novel low-profile external skeletal fixator for type IIIB open tibial fractures: a biomechanical and clinical pilot study. J Chin Med Assoc: JCMA. 2021;84(5):528-535.
    [Google Scholar]
  14. , , , , . The role of external fixation in the treatment of posttraumatic osteomyelitis. J Orthop Trauma. 1988;2(2):79-93.
    [Google Scholar]
  15. , , , . Use of the unilateral external fixator (monofixator) in fracture treatment: experience in 50 fractures. Unfallchirurg. 1994;97(10):511-517.
    [Google Scholar]
  16. , , , , , . Factors in pin tract infections. Orthopedics. 1991;14(3):305-308.
    [Google Scholar]
  17. , , , . The influence of intermittent micromovement upon the healing of experimental fractures. Orthopedics. 1984;7(3):481-484.
    [Google Scholar]
  18. , , , et al . The effects of external mechanical stimulation on the healing of diaphyseal osteotomies fixed by flexible external fixation. Clin BioMech. 1998;13(4-5):359-364.
    [Google Scholar]
  19. , . Physical and biological aspects of fracture healing with special reference to internal fixation. Clin Orthop Relat Res. 1979;138(138):175-196.
    [Google Scholar]
  20. , , , . The effect of movement on the holding power of screws in bone. Clin Orthop Relat Res. 1975;111(111):257-262.
    [Google Scholar]
  21. , . Mechanical factors influencing the holding power of screws in compact bone. J Bone Joint Surg. 1973;55(3):633-639.
    [Google Scholar]
  22. , . Metal implants and surface reactions. Injury. 1996;27(Supplement 3):S/C16-C22.
    [Google Scholar]
  23. , , , , , , . Influence of materials for fixation implants on local infection. An experimental study of steel versus titanium DCP in rabbits. Journal of Bone and Joint Surgery - Series B.. 1996;78(4):647-651.
    [Google Scholar]
  24. , , , , , . Tierexperimentelle Untersuchung zur Haftfestigkeit von Hydroxylapatit-beschichteten AO/ASIF-Schanz-Schrauben für den Fixateur externe [Animal experiment study of tensile strength of hydroxylapatite-coated AO/ASIF Schanz screws for external fixators] Unfallchirurg. 1994;97(8):391-398.
    [Google Scholar]
  25. , , , , , . External fixation pin: an in vitro general investigation. Orthopedics. 1987;10(11):1517.
    [Google Scholar]
  26. , , , , . Axial preload in external fixator half-pins: a preliminary mechanical study of an experimental bone anchorage system. Clin BioMech. 1999;14(1):69-73.
    [Google Scholar]
  27. , , , . Cortical bone reactions at the interface of external fixation half-pins under different loading conditions. J Trauma. 1993;35(5):776-785.
    [Google Scholar]
  28. , , , . Analysis of the external fixator pin-bone interface. Clin Orthop Relat Res. 1993;293:18-27.
    [Google Scholar]
Show Sections