Research Article
2025
:4;
100418
doi:
10.1016/j.jorep.2024.100418

Translate:

Ilizarov wire tensioning - The ring deflection phenomenon

Author affiliations
Department of Trauma and Orthopaedic Surgery, Queen Elizabeth University Hospital, Glasgow, United Kingdom
HealthPoint & Milton Park Medical Centres, Belgravia, Harare, Zimbabwe
School of Medicine, College of Medical Veterinary & Life Sciences, University of Glasgow, United Kingdom
Article History
Corresponding author(s)

⁎Corresponding author: P.R.M. Nicholas. patricknicholas1997@gmail.com

Article tools · PDF & citations

Abstract

Abstract

Gavril Ilizarov pioneered the use of fine wire circular ring fixators within Soviet Russia in the early 1950's and advocated a wire tension within the range of 980 N–1275 N. Ilizarov used a technique known as “Russian tensioning” whereby a wire is subtended around the fixation bolt in order to produce wire tension across the frame. The use of a mechanical tensioning device was conceived to elongate the wire and impart tension on the wire.

This biomechanical study compares these two techniques at achieving adequate tension and aims to quantify how initial wire tension is affected by the addition of a second, orthogonal wire.

In this in vitro series of experiments, a stabilised, testing rig utilised an initial wire passed across a 160mm stainless-steel ring. A Russian tensioning method and mechanical tensioning device was used. A calibrated tensiometer was used to record spoke deflection which was then converted to Newtons of tension.

When Russian tensioning, adequate tension was only achieved at 90° of bolt rotation. Notably the first wire tensioned did not achieve a suitable tension until the 2nd wire was added across all angles of bolt rotation.

Only at 110 and 130KgF was initial wire tension adequate. Following the addition of an orthogonal wire at both of these settings wire tension was maintained within the acceptable range.

Both methods of tensioning are an effective method of tensioning in Ilizarov constructs. A "ring deflection phenomenon" may cause an initial loss of tension but the addition of an orthogonal wire ensures adequate wire tension, provided there is 90° of bolt rotation with Russian tensioning or at least 110 KgF with a mechanical tensioner.

Keywords

  • Ilizarov
  • Biomechanics
  • Wire-tensioning
Disclaimer:
This article was originally published by Reed Elsevier India Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.
1

1 Introduction

The Ilizarov method of transosseous osteosynthesis based on Gavril Abramovich Ilizarov's modular frame is a highly effective method of managing otherwise challenging scenarios relating to trauma and its sequelae.1,2 Numerous studies to fully understand its biomechanics have been carried out since its inception in Siberia in the 1950's, both within and outside the former Soviet Union and present-day Russia.3–6 This is no surprise as a thorough understanding of Ilizarov's methods is required to successfully use an Ilizarov circular ring fixator.

A key concept of this method is the use of tensioned wires. These are applied perpendicular to the weight bearing axis of the bone to achieve fixator stiffness with minimal wire deflection and optimal interfragmentary micromotion. This mechanism is what makes this method superior to conventional methods at achieving bony union.3,4,7–9 There is a set limit as to how much interfragmentary micromotion is desirable. There have been a number of research papers investigating wire tensions achieved by hand and by a mechanical tensioning device. An optimal wire tension ranging from 980 N (100Kg) to 1275 N (130Kg) is required to achieve sufficient tension to allow early joint motion and weight bearing.1,3,4,10,11

Ilizarov employed a wire tensioning method known as “Russian tensioning”. This method involves the use of a simple pair of spanners whereby the wire is subtended around the fixation bolt to tension the wire across the ring. However, this may be difficult to replicate as it is based purely on feel, trial and error, or the anecdotal reports of experienced users.1,12,13 (Fig. 1) The innovation of a graduated, calibrated spring tensioning device was developed to aid surgeons when attempting to achieve the desired wire tension.1 It was subsequently noted that wire tension loss of up to 30% occurs in the typically stainless-steel wires used in most cases.4,8 A plethora of reasons have been suggested to contribute to this phenomenon. They include, but are not limited to, the type of wire fixation method to the rings, the propensity for ductile stainless-steel wires to yield at this interface and the ability of the wires to slip due to insufficient fixation bolt torque.2,5,14,15

‘Na Raskrut’ method of Russian tensioning method showing spanners being used to tension the wire around a fixation bolt. The bolt and nut rotate in opposite directions to achieve tension, initially the slotted bolt to elongate the wire, then the nut to tighten.
Fig. 1 ‘Na Raskrut’ method of Russian tensioning method showing spanners being used to tension the wire around a fixation bolt. The bolt and nut rotate in opposite directions to achieve tension, initially the slotted bolt to elongate the wire, then the nut to tighten.

The mechanism of the calibrated, spring-loaded tensioner device involves grasping the wire and then using clockwise rotation of the tensioner to elongate the wire in a linear fashion to achieve the desired force. The tensioner has a footprint at its base to allow resting upon the frame whilst tensioning. The user can set the tensioner to the required Kilogram-Force (KgF) value which in turn can then be converted to the corresponding tension. This study's design is based on the measurement of the tension generated and lost in a single wire model mounted on a half-ring rigidly fixed to another surface. It was hypothesised that the tension achieved with an initial wire would be impacted by subsequent wires added to a ring fixator.

There have been a number of research papers investigating the tensions achieved by hand and by the tensioning device.

To our knowledge, the effects of a second orthogonal wire on a full ring used in practice on the initial (first) wire tension have not been studied.

In this biomechanical study, this study aims to quantify the size of the effect of a second tensioned wire on the initial (first) wire tension and how this compares between both tensioning methods. In order to do this, we first require to quantify the tension achieved when one wire has been passed across the ring.

2

2 Methods

2.1

2.1 Initial tensioning

A wire-tensioning setup was constructed using a 160mm stainless-steel ring. The ring was securely held by a single clamp. A single, 1.8mm stainless steel wire (grade 316L) was then passed across the ring construct. Our first set of experiments involved Russian tensioning and our second set of experiments involved the use of mechanical tensioners. Wires that were Russian tensioned were set to predetermined angles of 45°, 70° and 90° using a goniometer for accuracy (Fig. 2). To evaluate the mechanical tensioning technique, a new and calibrated tensioner (IdealMed, Chester, United Kingdom) was employed as per the manufacturer's guidance to load the wire with predetermined force (90KgF, 110KgF, 130KgF) (Fig. 3). Steel fixation bolts were used to attach the wire to the ring. Following addition of the first wire, a new tensiometer (DT Swiss, Biel, Switzerland) was then used to measure deflection across the wire (Fig. 4). A calibration curve was created to accurately calculate measured tension (N) from deflection (mm). Each data point was triplicated and standard deviations noted.

Russian tensioning with a goniometer used to accurately tension to angles of 45°, 70° and 90° (Shown left to right).
Fig. 2 Russian tensioning with a goniometer used to accurately tension to angles of 45°, 70° and 90° (Shown left to right).
Calibrated, spring-loaded tensioner used for Italian tensioning.
Fig. 3 Calibrated, spring-loaded tensioner used for Italian tensioning.
DT SWISS tensiometer used to measure spoke tension across the wire.
Fig. 4 DT SWISS tensiometer used to measure spoke tension across the wire.
2.2

2.2 Addition of an orthogonal wire

Following tensioning of the first wire, a second, orthogonal wire was then added to our construct using the same type of wire and method of tensioning as the first wire. The tension in the second wire was recorded. Following this, the tension across the first wire was then re-measured. Each data point was again triplicated and standard deviations noted.

2.3

2.3 Statistics

Kolmogorov-Smirnov test was performed. T-tests were used to compare mean tensions. (GraphPad Prism, v6 for Macintosh). A p-value of <0.05 was considered statistically significant.

2.4

2.4 Data collection

Results were collated by the same clinician throughout. This was under the close supervision of the 2 senior surgeons of this paper, both of whom provided extensive teaching of both tensioning techniques to the clinician collating the data.

3

3 Results

Russian tensioning achieved a mean tension of 1046.7 N at 90° of bolt rotation which was within the target tension range of between 980 and 1275 N (Table 1). Notably the first wire tensioned did not achieve a suitable tension until the 2nd wire was added across all angles of bolt rotation. At 45° and 70° target tension was not achieved, however the addition of an orthogonal wire did increase the tension upon the first wire, albeit not to the target tension as proposed by Ilizarov.

Table 1 Mean wire tensions for Russian tensioning method (SD). Results in Newtons (N).
Russian Tensioning Angle (°) Initial wire 1 Tension (SD) Wire 2 Tension (SD) Subsequent Wire 1 Tension (SD) P-Value
45 109.7 (16.2) 430.0 (43.6) 373.0 (11.5) <0.0001
70 121.0 (46.9) 401.7 (44.8) 455.0 (74.0) 0.002
90 98.6 (13.7) 1238.3 (245.1) 1046.7 (353.7) 0.013

Evaluation of the mechanical tensioner revealed that the target tension range of 980–1275 N was also not achieved until the addition of an orthogonal wire at the 90KgF setting. The initial wire tension was adequate at 110 and 130KgF (Table 2). Following the addition of an orthogonal wire at both of these settings wire tension was maintained within the acceptable range. At 110KgF, the tension increased in the initial wire, whereas at 130KgF the initial wire tension decreased. The mechanical tensioner appeared to underestimate tension at lower settings, however at higher settings (>130KgF) tension is possibly overestimated (Fig. 5).

Table 2 Mean wire tensions for Mechanical Tensioner method. Results in Newtons (N).
Tensioner Setting (KgF) Initial wire 1 Tension (SD) Wire 2 Tension (SD) Subsequent Wire 1 Tension (SD) P-Value
90 638.55 (28.1) 836.7 (46.1) 790.46 (95.4) 0.048
110 1084.86 (53.1) 1201.31 (33.3) 1122.22 (66.9) 0.285
130 1483.95 (104.7) 1141.33 (205) 1380.15 (111.0) 0.195
Difference between proposed tension on device and measured tension achieved the Italian tensioner at different settings.
Fig. 5 Difference between proposed tension on device and measured tension achieved the Italian tensioner at different settings.
4

4 Discussion

This work has demonstrated that both methods of tensioning may achieve adequate tension upon a stainless-steel wire. In our experience, the Russian method is a quick and effective method of achieving tension however has limitations, such as the difficulty should a wire require retensioning.

In order to ensure the accuracy of the mechanical tensioners is maintained, these should be cleaned and lubricated at regular intervals.

When Russian tensioning, only at an angle of 90° can a sufficient tension be achieved across the initial wire to the suggested range of between 980 and 1275 N.3,4,10,11 These relatively low volumes of tension may be caused by initial deformation of the ring as the first wire is passed. This is lower than we had expected and when compared to Davidson et al. published work, they produced much higher tensions following the passing of a wire which was tensioned using the Russian technique.16 When an orthogonal wire is then added the wire tension of the initial wire increases significantly. This may be explained by tension redistribution across the ring. This acts to counterbalance the forces achieved, causing an increase in the initial wire tension. When applying the Ilizarov circular ring fixator in clinical practice, this research shows that surgeons first must endeavour to achieve 90° of bolt rotation when Russian tensioning and secondly recognise the importance of an orthogonal wire in increasing the tension across the initial wire.

As recurrent load cycles are being exerted on an Ilizarov frame, when a patient bears weight through it, the construct wires will begin to fatigue over time or slip through the bolts. This often results in the surgeon requiring to re-tension the wires in the outpatient setting. As wires are often cut following tensioning at the end of the operative procedure, Russian tensioning using spanners can be a much more accessible way of retensioning.

When the mechanical tensioner was employed, our results were more varied. It is important to note firstly that the amount of tension across the initial wire is significantly higher globally than what was demonstrated in the Russian tensioning data set. The tension across the initial wire and the resultant tension after orthogonal wire placement demonstrated satisfactory results with adequate tension being achieved at both 110 and 130KgF. In clinical practice, similarly to the Russian tensioning group at 90° of bolt rotation, it is important to note that by setting the tensioner to those markers, we can assume that adequate tension is going to be achieved following the addition of an orthogonal wire across the Ilizarov frame.

Our results demonstrate that it may be a less predictable method for producing wire tension than Russian tensioning. One explanation of this could be the condition of the tensioner used.5,16 In our study a single tensioner was used, albeit new, it was used for a large data set and therefore may be subject to wear and therefore inaccuracies. Furthermore, the mechanical tensioning device does not allow direct measurement of tension. It estimates load based upon predetermined elongation which is based upon a given wire width and material.

4.1

4.1 Limitations

This study is limited by the use of a single stainless-steel ring within the series of experiments. This may have given rise to fatiguing of the construct. A progressive reduction in load values was not however observed, and the use of a single ring for more than 100 wires pales into insignificance when compared to the ultimate strength and cyclical loading resistance the rings are engineered to withstand.

Our ring was held using a rigid clamp device, it would be beneficial for future studies to add a ring and connect this using threaded rods in order to simulate with more accuracy how this would be achieved in real life, in case this has a bearing on the results achieved.

In addition to this, our study had optimal conditions for the placement of an orthogonal wire. This situation is rare given the constrictions of anatomical corridors.

5

5 Conclusion

Russian tensioning appears to be a fast and effective method for tensioning Ilizarov constructs, however a minimum angle of 90° should be sought to achieve adequate tension. We have demonstrated that the “ring deflection phenomenon” may account for an initial loss of tension, however the addition of an orthogonal wire recovers this in construct stability. The mechanical tensioners, although able to provide sufficient tensions at higher settings may have increased unpredictability. However, they do provide a reasonable alternative to Russian tensioning in providing tension across a wire.

Clinically, our research demonstrates that by adding an orthogonal wire will ensure adequate wire tension as long as 90° of bolt rotation occurs if Russian tensioning or that at least 110KgF is used if a mechanical tensioner is used.

Further exploration of the deflection phenomenon (and its subsequent resolution with a second wire at various tangents) is required to give the operating surgeon an estimation of the ultimate tension achieved with fine wires in a frame construct. This research should explore the impact of deflection across varying ring diameters (including incomplete rings and foot plates) and differing wire subtending angles. This will give the user a more complete estimation of the ideal tensioned construct when faced with the anatomical limitations which challenge us in each individual case.

Ethics

No Ethical approval was required for this study.

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors however institutional funding has been received for research and education from Zimmer, Stryker, DePuy Synthes, UKRI, Hospital Innovations and Idealmed.

No patient consent was required for this research paper.

CRediT authorship contribution statement

P.R.M. Nicholas: was involved in data collection and analysis, statistical analysis, manuscript preparation and drafting. J. Mthethwa: was involved in manuscript preparation and drafting. B. Jamal: was involved in data analysis and manuscript drafting. D.W. Shields: was involved in concept design, statistical and data analysis and manuscript drafting.

References

  1. , , , , , , . Wire tension in the Ilizarov system: accuracy of the wire-tensioning device. Engineering in Medicine Proc. IMechE. 2003;219
    [Google Scholar]
  2. , , , , . The Ilizarov method in nonunion, malunion and infection of fractures. J Bone Joint Surg. 1997;79(2):273-279.
    [Google Scholar]
  3. , , , , . The effect of clamping a tensioned wire: implications of the Ilizarov external fixation system. Proc. Inst. Mech Engrs. Part H: J.Engineering in Medicine. 2003;217:91-98.
    [Google Scholar]
  4. , , , , . The biomechanics of wire fixation in the Ilizarov system. Injury. Int. J. Care Injured. 2003;34:155-157.
    [Google Scholar]
  5. , , , , . Mechanical stress on tensioned wires at direct and indirect loading: a biomechanical study on the Ilizarov external fixator. Int. J. Care Injured. 2011;42:1107-1111.
    [Google Scholar]
  6. , , , , . Yielding of the clamped-wire system in the Ilizarov external fixator. Engineering in Medicine Proc. IMechE. 2003;217
    [Google Scholar]
  7. , , . Effect of wire tension on rate of weight bearing in Ilizarov apparatus. Am J Biomed Sci & Res. 2019;3(4)
    [Google Scholar]
  8. , , . Mechanical performance of external fixators with wires for the treatment of bone fractures. Part II: wire tensioning and Slippage. J. Biomech. Engng. 1993;115:37-42.
    [Google Scholar]
  9. , , , , , . Cyclical micromovement and fracture healing. J Bone Joint Surg. 1995;77(4):645-648.
    [Google Scholar]
  10. , , , , , . Wire tensioning with integrated load-cell in the Ilizarov orthopedic external fixation system. Mech Mach Theor. 2014;79:109-123.
    [Google Scholar]
  11. , , , et al . The effects of mechanical stimulation on bone healing of diaphyseal osteotomies fixed by external fixation. J Biomechanics. 1998;13:359-364.
    [Google Scholar]
  12. , , , , , , . Mechanisches Verhalten von Ilizarov-Ringfixateuren. Unfallchirurg.. 2000;103:839-845.
    [Google Scholar]
  13. , , . The Original Ilizarov System: General Surgical Technique Brochure. 1998:6-7.
    [Google Scholar]
  14. , , , . Rapid pre-tension loss in the Ilizarov external fixator: an in vitro study. Journal Acta Orthopaedica. Jul 2009;78(2007 - Issue 5):654-660.
    [Google Scholar]
  15. , , . Mechanical considerations in using tensioned wires in a transosseous external fixator system. Clin Orthop. 1992;280:23-29.
    [Google Scholar]
  16. , , , , . Ilizarov wire tensioning and holding methods: a biomechanical study. Injury. 2003 Feb;34(2):151-154.